QA Checking of unit dose preparation
Steps involved:
1. Scanning a pick list of products to repackage against the stock containers selected for repackaging to make sure the proper stock containers were selected.
(
NDC to NDC # matching)
2. Scan the UD label’s NDC # bar code and visually compare the drug information on the RxScan Ultra UD Check’s
screen to the drug information printed on the UD label. This verifies that the correct bar code has been placed on the label.
(NDC # bar code scanning to view the matching drug information)
3. Scan the UD label NDC # bar code and compare it to a scan of the stock container’s NDC # bar code. This is to assure that the proper product for the
UD label was selected before repackaging.
(NDC to NDC # matching)
4. Provide a mechanism for the pharmacist to double-check the final product, i.e. after the repackaging is completed they can verify that the right label with the right drug were prepared.
(NDC to NDC # matching & scanning to view a picture of the product and the drug information that should be printed on the label)
|